The miniDIA deionizers line has been designed to be handy, compact and economic, ready to adapt to plant situations with low water flow.
Thanks to different configurations, the miniDIA deionizers can supply 10 to 30 liters / hour of treated water according to the CLSI standard.
PLUS-T versions include a treated water storage tank with a capacity up to 50 liters.
Deionized water production up to 30 l/h.
CLSI Standard water, with a typical conductivity of 0.1 µS/cm
Fully automatic versions, with less interaction with user.
With automatic bypass, the DIA guarantees the flow of deionized water in all conditions *.
Optimization of components to ensure reduced dimensions and easy installation.
Very easy-to-use interface and disposables replacement wizard.
Technical description
Denomination: Mixed bed of ion-exchange resins
Brand: Quality Invents
Model: ION tank
Code: Q-1000-10
Manufacturer: Quality Invents Srl – Italy
Product expires: 12 months after production
Type of product: Disposable consumable
Technical description
Denomination: Advanced Triple Action Filter
Brand: Quality Invents
Model: μQ
Code: Q-1000-25
Manufacturer: Quality Invents Srl
Nation of production: Italy
Expiration date: No
Type of product: Single-use consumable
1. Check that the power cable is correctly connected to the DIA and to the electric socket.
2. Check that the socket is working.
3. DIA switch must point to “I/ON”.
4. Unplug the power cable from DIA power block and check the fuses; if necessary, replace them with new ones of the same type and amperage. Repeat the procedure from point 1). If the system does not switch on…
5. Power off the DIA turning the switch to “0/OFF”, turn the manual by-pass valve to “by-pass” (to the left) and immediately call the Customer Service.
1. Turn the manual by-pass valve to “by-pass” (to the left). Leave the DIA on, so that you notice when the power comes back again.
2. When the power supply returns, turn the by-pass valve to the normal working position (to the right).
1. The deionizer must not be installed on benches, furniture or other devices.
2. The front panel and the back panel must be close.
3. If, as well as the noise, you notice that the device is overheating, power off the DIA by turning the switch to “0/OFF”, turn the manual by-pass valve to “by-pass” (to the left) and call the Customer Service.
1. Switch off the deionizer and activate the manual By-pass mode.
2. In these conditions the IONtank will last for a much shorter time.
3. When you have the new consumables replace it.
1. Power off the DIA turning the switch to “0/OFF”, turn the manual By-pass valve to “by-pass” (to the left), continue to work as usual and call the Customer Service.
1. Inspect the inside of cylinder housing as best as you can (plus version)
2. If there is a serious leak, switch off the miniDIA and wait for the Customer service to arrive. When the miniDIA by turning the switch “OFF”, turn the in-feed tap to “By-Pass” and call immediately Customer service.
3. Never leave the deionizer ON with the water in-feed tap open, especially with AB3 alarm (if enabled) or if there are any water leak.
1. More water is being drawn off than the deionizer can produce. Contact our Customer service.
2. The prefilters are blocked and are limiting the module’s production, try to replace them.
3. Have you ever received any AH1 messages, even occasionally? Check previous warnings in the register.
4. Call Costumer Service.
1. Your average water requirement has increased
2. The miniDIA has remained in By-Pass mode for too long (PLUS version), manual or automatic.
3. The total salinity of the raw water has increased. To measure this value, call our technical customer service.
4. Call Costumer Service.
1. Check the right housing of miniION cartridge in miniDIA version
2. Check the connections on the IONtank in PLUS version.
3. Be sure that delivery pipe is not bent
4. Check if you put the miniDIA in By-Pass (PLUS version) the water is begin delivered: if not, this means that the pump unit is damaged; otherwise there must be a problem with IONtank connections.
5. The pump is damage. Call costumer service.
The system has signalled there is not inflow of water.
The deionizer has a reserve of water for the osmosis process.Product water supply is guaranteed by the water reserve.
1- Make sure that the in feed tap is open.
2- If the raw water inflow is equipped with a prefilter unit, check that the prefilters are not obstructed.
3- Check that the inlet pipe is not blocked
4- Check that the manual feed-tap is not in by-pass (Plus version) or close position.
5- If the problem persist call Customer Service.
The system has signalled that there is no water inflow.
The minimum level sensor has not been covered in 20 seconds. The deionizer has a reserve of water for the osmosis process.
Product water supply is guaranteed by the water reserve.
See solution message AI1
The system has signalled that there is no water inflow.
The deionizer has no water reserve for osmosis process.
Product water supply is guaranteed by reserve water.
See solution message AI1
The system has calculated that the time taken to fill the tank is longer than the period indicated in the specifications. The capacity is not sufficient to guarantee the nominal productive inflow of the system. If the problem persists, the alarm sounds every 12 hours even if it has been switched off. The service is guaranteed depending on the average inflow of mains water
1-Check raw water supply line, feed tap, external prefilter, water pressure.
2-Replace miniDIA prefilters
3-If the problem persist, call customer service.
The system has calculated that the tank filling time is longer than the period indicated in the specifications. The capacity is not sufficient to guarantee the nominal product inflow of the system.
The system warns, that is necessary replace the prefilters. You have 24 solar hours to replace the prefilters.
Water supply available.
1- Replace the old pre-filter with a new one (see section 11.1-11.2)
2- Don’t turn off or stop miniDIA while you are doing the prefilter replacement.
23 hours have gone since the miniDIA asked you to replace the prefilters AP1. The system is asking for immediately prefilters replacement.
The miniDIA is in By-Pass (only PLUS version) to not stress the osmosis system.
In one hour the system will stop. Product water supply not available.
1- See solution message AP1.
24 hours have gone since the miniDIA asked you to replace the pre-filter AP1. The system has not feel the pre-filter replacement, the miniDIA now is stopped. Product water supply not available.
1- See solution message AP1.
2- If you still have problem, please call Customer Service.
The system has noticed an illogical combination of level sensor group in the treated water tank (TW).
In the miniDIA version water supply is not available.
In the miniDIA PLUS version the system is positioned automatically in By-Pass, water supply is available.
1- Call Customer Service.
The system has noticed an illogical combination of level sensor in the raw water tank (RW).
In the miniDIA version water supply is not available.
In the miniDIA PLUS version the system is positioned automatically in By-Pass, water supply is available.
1- Call Customer Service.
The leakage sensor (water leakage, if enabled) has noticed a water leakage.
In the miniDIA version water supply is not available.
In the miniDIA PLUS version the system is positioned automatically in By-Pass, water supply is available.
1- Call immediately Customer Service.
The treated water tank TW is empty. The machine is producing water to restore the correct level to guarantee service continuity.
In the miniDIA version water supply is not available.
In the miniDIA PLUS version the system is positioned automatically in By-Pass, water supply is available.
1- Call Customer Service.
The system has noticed an anomaly in the conductivity reading cell.
1- Stop the alarm pressing “ESC” and continue normal operation
2- Call Customer Service.
The system has noticed that product water conductivity has reached the maximum value programmed (usually 0.8 µS/cm).
In miniDIA version, replace the exhausted miniION cartridge (section 11.3 of the manual) and condition it (section 11.4 of the manual).
In miniDIA PLUS version, replace exhausted IONtank (section 11.4 of the manual) and condition it ( section 11.5 of the manual)
1- In the miniDIA version replace exhausted cartridge and condition it.
2- In the miniDIA PLUS version replace the exhausted IONtank with a new one and condition it.
Do not switch off the miniDIA refer to this manual and perform the detailed operation.
The system has noticed that the osmotic module is producting less than the rated value.
1- If the miniDIA has not been working for some days wait a few hours before calling the Customer Service, the alarm may deactivate itself.
2- If the alarm sounds again or the miniDIA has been not working, call the Customer Service.
3- Switch off the alarm and continue normal operation.
Buffer battery level of charge is too low or there are problems with the internal clock. The machine work sas usual.
1- Call Customer Service.
(@15 °C, 800ppm NaCl, pH 8,3 e SDI 5)
wdt_ID | Colonna 1 | Colonna 2 | Valore |
---|---|---|---|
1 | Salt rejection | % | 95 |
8 | Applied pressure | bar: | 10 max |
9 | Permeate flow | l/h: | 10-30 * |
10 | Disposable replacement | in Stand By |
CLSI Standard
wdt_ID | Colonna 1 | Colonna 2 | Valore |
---|---|---|---|
1 | Resistivity | MΩ.cm: | 10 |
2 | Bacteria | CFU/ml: | < 10 * |
8 | TOC (Total Organic Carbon) | ppb: | < 500 (with 185nm UV Lamp) |
9 | Nominal Flow Rate | l/min: | 2 (120 l/h) |
10 | Pressure | bar: | 0,9 - 3,0 adjustable |
Via E.Ferrari, 23 - 20010 Arluno (Mi) - Italy || Tel. 0039 029017224 Email help.dia@quality-group.eu
Maybachstraße 16 - 67269 Grünstadt - Deutschland || Tel. 0049 6359936943 Email info@ass-gmbh.com
5 rue Gutenberg - ZI la mariniere - 91070 Bondoufle FRANCE || Tel. 0033 0164972780 Email contact@asd-labo.com
Calle Chile, 10, 28290 Matas-Pinar-Monte Rozas (Las), Madrid, Spagna || Tel. 0034 916300085 Email comercial@aadlab.com
401, 402ho, Seo Young Venice Square 61, Seouldaehak-ro 278 beon-gil, Siheung-si, Gyeonggi-do, Korea 15011 || Tel. +82-31-499-6917 Email hansol@hansolmedi.com
Golden time business centre office M02-06C Dubai United Arab Emirates || Tel. +971 56 2263058 Email: ahmed.ghanem@agilelabme.com